Бренд:NANBEI
Модель:NB988-4
Премьера продукта:
Технические параметры:
Thermal Block
|
||||
Block formats
|
96-well
|
|||
Sample volume
|
5-50ul
|
|||
Consumables
|
0.1ml 96-well unskirted Plates; 0.1ml 8-strip tubes (Optical flat cap)
|
|||
Max. heating rate
|
4.2℃/s
|
|||
Max. Cooling rate
|
4.2℃/s
|
|||
Temperature range
|
4.0-99.0℃
|
|||
Temperature accuracy
|
±0.1℃
|
|||
Temperature uniformity
|
±0.3℃
|
|||
Temperature steps
|
1-99
|
|||
Heat Lid
|
||||
Temperature range
|
100-120℃
|
|||
Control
|
Automatic temperature and pressure setting
|
|||
Optics
|
|
|||
Excitation
|
4LEDS
|
|||
Excitation range
|
470-630nm
|
|||
Detector
|
Photodiode
|
|||
Detection range
|
525-670nm
|
|||
Detection channels
|
|
|||
Channel
|
Excitation(nm)
|
Emission(nm)
|
Pre.Calibrated Dyes
|
|
1
|
470
|
525
|
FAM,SYBR CREENI
|
|
2
|
523
|
564
|
VIC,HEX,TET,JOE
|
|
3
|
571
|
621
|
ROX,TEXAS-RED
|
|
4
|
630
|
670
|
Cy5
|
|
Multiplex
|
Up to 4 targets
|
|||
Detection range
|
1-10^10copies
|
|||
SOFTWARE
|
||||
Режимы анализа
|
Абсолютное количественное определение, относительное количественное определение, анализ кривой плавления, плавление с высоким разрешением
|
|||
PERFORMANCE
|
||||
Воспроизводимость Коэффициент корреляции точности нуклеиновых кислот
|
CV≤2.1%
R≥0.997
|
|||
SYSTEM
|
||||
Operating systems
|
Windows XP,Windows 7
|
|||
Communication
|
USB or Ethernet
|
|||
Power usage
|
AC220*(1±10%)V,50*(1±2%)Hz
|
|||
Dimensions
|
540mm*300mm*480mm
|
|||
Weight
|